Compare buying individual styli vs. kits for CMM probes: ordering, delivery, installation, inventory, and decision criteria for M2/M3/HP-S-X1C.
Should I buy a stylus kit or individual styli?
This is the most frequently asked question in the metrology room. Most of the people asking have just taken over a CMM or changed their probe head, and are holding a catalogue in their hands. When they reach the “Stylus Kit” pages, they stop: the kit contains straight styli, extension bars, star styli and adaptors, which looks comprehensive. But when they actually go to use it, they find the stylus tip diameter does not match, the length is insufficient, and the thread may not fit either. So they start flipping through the individual stylus specification table again, checking L, D, S and EWL line by line, and the more they look, the more confused they become.
The root of the confusion lies not in the styli themselves, but in the fact that kits and individual styli follow two different procurement logics. Buying individually is driven by the features to be measured: first identify the probe head model, then the thread, then the stylus tip diameter and length, with each step following the specific workpiece. Buying a kit is packaged around “common scenarios”: the manufacturer bundles a set of frequently used tip diameters and lengths together, saving the trouble of ordering each item individually. But it also means you carry inventory and bookkeeping costs for the few styli you never use. Both logics are valid, but they suit different situations — and the most common mistake on the shop floor is applying one logic to the other scenario.
This article aims to resolve the question of whether to buy a kit or individual styli.
First, let’s clarify the real differences between buying individual probes and buying a complete set across the four stages: ordering, delivery, installation, and inventory. We won’t discuss price—only process and risk. Then we’ll present the three stylus kits listed in the official catalogue: 03969075 for M2, 03969040 for M3, and M00-694-008-000 for HP-S-X1C, covering their intended use and typical contents, based on the official data sheets. Next, we’ll break down the four types of components usually included in a kit—straight styli, extension bars, star styli, and magnetic mounts—explaining what each is for and why the manufacturer bundles them together.
Next, we’ll cover the decision criteria. Buying a complete set is more convenient when: installing a new machine, replenishing a batch of common sizes in the spare parts inventory, or when the shop floor mainly uses small- to medium-diameter straight styli with a steady demand for extension bars. Individual purchase is required when: the stylus ball diameter is non-standard, the length must be matched to a specific EWL, or the probe is a scanning type and the kit lacks the corresponding mount and adaptor. Finally, we’ll address kit maintenance—how to replace missing parts, how to record mixed kits, and how to set up a tracking ledger—because the real challenges with a kit usually appear not on the day you buy it, but six months into use.
After reading this article, you should be able to take your probe model and a workpiece list and decide whether to buy individual styli or a kit, or a mix of both. You will also know which pages to look at in the catalogue and in what order, without being put off by a 72-page specification table. All part numbers, page numbers, and parameters mentioned here are taken from the official catalogue and data sheets; we do not speculate on combinations not listed in the tables.
Buying individual styli and buying a kit are two different decision-making approaches. The rule of thumb: if the measured features in your task are consistent in size, depth, and orientation, individual styli are more economical; if you run multiple product types, small batches, frequent changeovers, or have not yet set up a stylus inventory on the shop floor, a kit is more convenient. The decision is not based on price but on changeover frequency and the maturity of your on-site management.
Buying individual styli is done on a per-task basis. For a fixed Ø 1 mm bore with constant depth and orientation, a straight stylus such as the M2 Stylus (R-1-SS-0.7-L10) plus an extension bar of the appropriate length is sufficient. The advantage is that specifications are fully controllable: select line by line from the official specification table, and check the seven parameters—part number, material, L (overall length), D (ball diameter), S (shank diameter), EWL (effective working length), and W (weight)—against the workpiece drawing. EWL is measured from the centre of the stylus ball to the point where the stylus shank would interfere with the feature being measured, whereas overall length L is measured from the mounting face to the ball centre. These two are not the same and must be considered together. The trade-off is a longer replacement chain: a broken stylus triggers a purchase, and a changeover triggers another. Without a proper inventory record, it is easy to end up with unclear specifications and unclear replacement requirements. Therefore, buying individual styli usually requires two prerequisites: stable tasks and someone maintaining the inventory ledger.
Complete sets are packaged by application scenario. The three sets in the official catalogue correspond to three scenarios: 03969075 is the M2 stylus kit 1, containing straight styli of various ball diameters and extension bars in lengths of 5/10/20/30; 03969040 is the M3 stylus kit 2, containing an M3 straight stylus and an extension bar; M00-694-008-000 is the M3 stylus kit (for HP-S-X1C), containing a suction cup and a star stylus holder, and is a dedicated scanning kit. The advantage of buying a complete set is that it is ready to use out of the box. When new equipment arrives, new production lines are being commissioned, or outsourced teams are equipped with general-purpose styli, there is no need to check each item individually. Extension bars, which are easy to miss, are already included in kit 1 in four lengths: 5/10/20/30. Another hidden benefit is a unified inventory starting point: one kit has one part number, so warehousing, requisition and stocktaking are all done by kit. Reporting the kit number for replacement parts is less error-prone than listing more than ten specification numbers item by item.
For maintenance, the challenge with buying individual styli is identifying missing parts: when one breaks, you must be able to state “M2 Stylus (R-1-SS-0.7-L10)” immediately so the replacement can be ordered correctly. If the inventory record only says “Ø1 stylus,” the ball material, rod material, rod diameter, and overall length are all guesswork, and the replacement may not fit or the EWL may be wrong. Therefore, when buying individually, keep a full inventory record with all seven fields from the official specification table. The challenge with buying kits is mixing: styli with different ball diameters look similar, and it is easy to mix them up when putting them away after use. Mixing across groups is worse—an M2 stylus placed in an M3 group will not thread correctly. So store them by group and record in the inventory which stylus belongs to which group. When a stylus is missing within a group, buying a single replacement is faster, while replacing the whole group is simpler; the choice depends on your management preference, but the replacement must be ordered by the original part number within the group, not described from memory.
Scenario: a workshop receives a new part requiring measurement of several Ø2 mm bore holes, a deep cavity, and a groove that must be approached from the side. Judgment order: first identify the probe/probe head model to determine whether the thread is M2 or M3 (the thread follows the probe, not the machine brand); then assess the ball diameter and length requirements for the three feature types; finally, check whether the existing stylus can cover them. If only two or three stylus tips are missing, buy them individually; if the gap is large or there is no systematic stock, buy a complete kit for the corresponding thread and include extension bars. The dividing line is not how many stylus tips you buy, but whether the gap is point-like or systemic.
Two general reminders: first, use the shortest stylus that meets the application and minimise the number of joints — every connection between stylus and extension bar adds a potential bending point, so keep this in mind during on-site setup. Second, match the stylus ball material to the workpiece material: ruby suits most applications; for heavy-duty scanning of aluminium parts, where adhesive wear is a risk, silicon nitride is recommended; for heavy-duty scanning of cast iron, where abrasive wear is a risk, zirconia is recommended. When buying individual stylus tips, it is easy to focus only on ball diameter and overlook the material; buying a complete kit reduces this risk because the material is consistent within the set. For the exact number of stylus tips and specifications in each kit, refer to the official data sheet in the catalogue layout and verify each item before ordering.
In the Hexagon Styli & Accessories catalogue (Chinese edition, 72 pages), the stylus kit (KIT) section lists only three part numbers: 03969075, 03969040 and M00-694-008-000. These three kits are not interchangeable alternatives at the same level; they are combinations for different thread sizes and different probe systems. When selecting a kit, do not simply compare which one contains more items — first identify the probe type on your machine.
| Kit | Part number | Description | Section |
|---|---|---|---|
| M2 stylus kit 1 | 03969075 | Includes straight styli with various ball diameters plus extension bars in lengths 5/10/20/30 | Part 1, M2 section (stylus kits, page 21) |
| M3 stylus kit 2 | 03969040 | Includes M3 straight styli and extension bars | Part 1, M3 section (stylus kits, page 30) |
| M3 stylus kit (for HP-S-X1C) | M00-694-008-000 | Scanning-specific kit with suction cups and star stylus holder | Part 2, M3 section (stylus kits, page 47) |
The catalogue is split into two parts: Part 1 covers “for touch-trigger probes” and Part 2 covers “for scanning probes”. Kits 03969075 and 03969040 are in Part 1; M00-694-008-000 is in Part 2. This division is the first filter: whether your machine uses a touch-trigger or a scanning probe determines which part you should look at.
M2 is the most common thread size for trigger probes. In the catalogue, the M2 section runs from page 14 to page 21, with dedicated subsections for styli, extension bars, accessories, and stylus kits — the most complete set of supporting information among the three kits.
The 03969075 kit consists of “straight styli with various ball diameters + extension bars in lengths of 5/10/20/30”. The catalogue lists the ball diameter options and quantities per option in column format, without item-by-item counts, so the official data sheet should be used as the reference. However, the structure of “multiple ball diameters + four extension bar lengths” makes the intended application clear: it covers common bore, step, and plane measurements, and when deep holes or interference avoidance are involved, extension bars can be combined to reach the required length.
The extension bar lengths of 5/10/20/30 are the only ones explicitly listed among these three kits. These four lengths are generally sufficient for on-site assembly, but note that each joint adds a potential bending point. The official selection guideline repeatedly stresses: use the shortest stylus that meets the application and minimise the number of joints. So extension bars should be “just long enough”, not “the longer, the more flexible”.
03969040 is the stylus kit for M3 threads, consisting of “M3 straight styli and extension bars”. In the catalogue, the M3 section is on pages 26 to 30, with dedicated pages for styli, extension bars, accessories, and stylus kits.
The difference between M3 and M2 is primarily the thread specification. If it doesn’t fit, it doesn’t fit—there is no adapter option (adaptors are covered in a separate section of the catalogue, but they introduce additional joint faces and overhang; the specific adaptor part number should be confirmed against the official data sheet). So when choosing between 03969075 and 03969040, the first step is not to compare contents, but to check whether the thread on the probe or probe head is M2 or M3.
How to tell: first identify the probe or probe head model, then determine the thread. The thread follows the probe, not the machine brand—many domestic CMMs and some other brands use the same probe system on the probe head or controller, so knowing the machine brand does not tell you the thread. You need to check the probe nameplate or the equipment documentation.
M00-694-008-000 is the only one of the three kits listed in Part 2 (for scanning probes). The catalogue marks it as “suitable for HP-S-X1C”. In addition to styli, it includes a suction cup and a star stylus holder.
Suction cups and star stylus holders are common accessories for scanning probes. A star stylus (Star Stylus / 5-way) allows styli to be mounted in multiple directions on a single holder, so features in several directions can be measured in one setup, reducing stylus changes. A suction cup enables quick stylus changes without screwing each stylus individually. Page 44 of the catalogue has a dedicated section on “Suction cups and adaptors”, and page 47 covers stylus kits.
The order number format for this set is M00-xxx-xxx-xxx, which is a different numbering system from the 03969xxx used by the first two sets. This reflects the two official numbering schemes: trigger probes use numbers starting with 03969, while scanning probes use numbers starting with M00. Do not mix the two when looking up part numbers.
| Your probe type | Thread | Corresponding set |
|---|---|---|
| Trigger probe | M2 | 03969075 |
| Trigger probe | M3 | 03969040 |
| Scanning probe (HP-S-X1C, etc.) | M3 | M00-694-008-000 |
Example: In a workshop, a CMM is fitted with a trigger probe on an M2 thread, used daily for measuring small holes and step surfaces. In this case, 03969075 is the correct set, and the four extension bar lengths cover most deep-hole scenarios. If you move to a machine with a scanning probe head such as the HP-S-X1C, then you need M00-694-008-000, because it includes a suction cup and star stylus holder, which are intended for scanning applications.
Conversely, if the machine has an M3-thread trigger probe but you order 03969075, the stylus will not fit. If you order M00-694-008-000, the thread matches, but the suction cup and star stylus holder may not be needed for pure trigger probing, so you end up with extra accessories you do not use. The point is not that more content is better, but that the set must match your probe system to be practical.
On page 34 of the catalogue, there is a Renishaw ↔ Hexagon cross-reference table; page 67 has the ITP ↔ Hexagon table. If you only have a part number from one of the other brands, you can look up the equivalent in the table, but do not assume a match for combinations not listed. The scanning table is based on the ITP system and does not include Renishaw part numbers — do not treat an ITP number as a Renishaw number when searching.
Second, the exact number of styli in each of the three kits, and the ball and rod diameters for each stylus, are listed in columns in the catalogue and are not itemised individually; the official data sheet takes precedence. When writing a request, it is recommended to narrow down the scope using three criteria — thread type, probe type, and whether a suction cup or star stylus holder is needed — then verify the details against the data sheet. This is faster than searching directly by part number.
When you receive a stylus kit, what determines whether it can be used is often not the number of styli, but the accessories — extension bars, star stylus holders, and suction cups. These determine how many workpiece shapes the same set of straight styli can cover, and whether a single missing component can be replaced on its own.
Straight styli are the main body. In the Hexagon catalogue, the M2 stylus kit Kit 1 (part number 03969075) is officially described as “containing multiple straight styli with various ball diameters + extension bars in lengths of 5/10/20/30”, and the M3 stylus kit Kit 2 (part number 03969040) as “containing M3 straight styli and extension bars”. Both kits bundle straight styli with extension bars, which shows that a single straight stylus cannot cover deep holes, steps, or internal cavities on its own. The specification format is Thread - Type (Ball Material - Ball Diameter - Rod Material - Rod Diameter - L Total Length), for example M2-Stylus (R-1-SS-0.7-L10) means M2 thread, ruby ball Ø1 mm, stainless steel rod Ø0.7 mm, total length 10 mm; M3-Stylus (R-3-SS-1.5-L21) means M3 thread, Ø3 mm ruby ball, Ø1.5 mm stainless steel rod, total length 21 mm. This naming convention is only used to explain the pattern and cannot be extrapolated to all part numbers; where no combination is listed, refer to the official data sheet. Material codes: R = ruby ball, TC = tungsten carbide rod, SS = stainless steel rod, CE = ceramic; prefixes such as SV/TI/DA refer to special geometries (disc, joint, cube, etc.), not ball material. Ball material selection: ruby suits the vast majority of applications; for heavy-duty scanning of aluminium parts, where adhesive wear is a risk, silicon nitride is recommended; for heavy-duty scanning of cast iron, where abrasive wear is a risk, zirconia is recommended.
Extension bars carry the stylus tip to depths beyond the reach of a straight stylus, with lengths available in 5/10/20/30 mm steps (refer to the official data sheet). The trade-off is that each joint adds a potential bending point. The official selection principle is: use the shortest stylus that meets the application requirements and minimise the number of joints. First check whether a straight stylus can reach the feature directly; if not, add an extension bar. If a 10 mm bar works, do not use a 20 mm one. The thread of the extension bar must match that of the straight stylus: M2 stylus with M2 bar, M3 stylus with M3 bar. The thread standard follows the probe head, not the machine brand — many domestic CMMs and some Mitutoyo models use Renishaw-style probe heads. The correct order of selection is: identify the probe head model first → then determine the thread → then choose the stylus tip diameter and length.
The star stylus holder is the most easily underestimated component. The comparison table lists the M2-Star Stylus 5-way (R-2-SS-1.4-SP20/30) (part number 03969056) and the M2-Styli Holder 5-way (SS-L7.5) (part number 03969054). The former is a five-way star stylus with a ball, and the latter is a five-way star holder. When measuring multiple sides, internal cavity walls, and flange faces of box-type parts, rotating the probe head is inefficient and can introduce secondary clamping errors. With the star holder installed once, the probe can reach multiple directions without moving. The M00-694-008-000 scanning-specific kit (for HP-S-X1C) explicitly includes a suction cup and star holder, confirming that the star holder is a standard configuration in scanning applications. The ball centre position and effective working length (EWL) of each stylus on the star holder must be calibrated individually. EWL is measured from the stylus ball centre to the point where the stylus stem would interfere with the feature being measured; total length L is measured from the mounting face to the ball centre. Refer to the official data sheet for exact values.
Suction cups mainly appear in the scanning probe group. In Part 2 of the catalogue, the M3 thread section lists “Suction cup and adaptor” (page 44), and the M5 thread section lists “Suction cup and adaptor” (page 61). This item is not included in the trigger-type M2/M3 section — it is an interface component for quick stylus changing on scanning probes. Scanning measurement often requires switching between multiple stylus configurations within one program, and the suction cup enables this switching automatically through a mechanical interface, rather than by manual threading. M00-694-008-000 includes a suction cup for this purpose. Suction cups and adaptors are components with an “additional joint surface”; each joint adds potential bending points, overhang, and loosening points, so they should be avoided where possible. However, when a program genuinely requires stylus changes, the repeat positioning consistency provided by the suction cup cannot be achieved by manually threading. Refer to the official data sheet for the specific adaptor part number.
The roles of the four component types can be summarised as follows: the straight stylus is the main body, determining the stylus tip diameter and base length — the tip diameter is chosen according to the measured feature size, and the stem diameter according to rigidity requirements; the extension bar addresses deep holes/internal cavities, and should be kept as short as possible to minimise joints; the star stylus enables multi-directional measurement in a single setup, with each stylus calibrated individually for EWL; the suction cup is the quick-change interface for scanning probes, and the fewer joint surfaces, the better.
When you receive a stylus kit, start with three questions: How deep is the deepest hole in the workpiece? If a straight stylus plus probe head overhang cannot reach it, an extension bar is essential. How many directions need to be measured in one operation? If more than two directions share the same datum, a star stylus can save one setup. Does the program require stylus changes? If it is a scanning probe, the adaptor plate should be in the kit; trigger probes usually do not need one. Compare your kit against these three criteria and add only the missing category—there is no need to buy a whole new set. If all three categories are missing, buying a complete set is usually simpler. The exact number of styli in each of the three KITs in the catalogue is not listed item by item because of the column layout; refer to the official data sheet. Before ordering, list the minimum bore diameter, maximum depth, and measurement directions from the workpiece drawing and check them against the catalogue—this is far more reliable than counting styli.
A stylus kit is not simply a bundle of styli sold together; it is a pre-configured combination for a typical set of working conditions. For equipment engineers and purchasing staff, the decision to buy a complete set or individual styli comes down to three factors: how stable the on-site conditions are, how fast spare parts turn over, and how easy the inventory records are to manage. In the following situations, buying a complete set is usually more convenient.
1. New equipment has just been commissioned and stylus requirements are not yet fully defined
During the new machine commissioning phase, no one can predict which dimensions will need to be measured over the coming year. Bore diameters from Ø1 to Ø10 may all come up, and deep holes, steps, sidewalls, and chamfers each place different demands on stylus tip diameter and length. Relying on a single stylus at this stage can easily lead to awkward situations, such as discovering halfway through a measurement that the stylus is not long enough, or that the tip diameter is too large to enter the bore.
Kits like 03969075 (M2 stylus kit 1) include straight styli with multiple tip diameters, and when combined with extension bars in lengths of 5/10/20/30 mm, they cover a typical range from short overhang to deep-hole probing. 03969040 (M3 stylus kit 2) is similar, combining M3 straight styli with extension bars. For the exact number of styli in each kit and their individual specifications, the catalogue layout is column-based, so it is recommended to refer to the official data sheet and verify each item before ordering.
Buying a kit during the commissioning period means having the commonly used styli on hand from the start, avoiding a procurement cycle every time a new feature needs to be measured.
2. Multi-shift operation and shared machines require high spare-part turnover
In a three-shift workshop, stylus collisions and tip wear are common. If you buy styluses individually, a damaged stylus has to go through requisition, approval, delivery and warehousing, and the machine may sit idle waiting for the replacement. Buying a kit gives you a buffer pool on site: the common sizes are already in the kit, so when one breaks you can pull a spare from the kit and reorder without stopping production.
This is especially true for the M2 thread. The official catalogue lists it for the TP2, TP20, TP200 and REVO™ touch-trigger probes, which are widely fitted on production lines and need frequent stylus changes. Keeping a 03969075 kit in the tool cabinet is a more reliable way to make sure a stylus is always on hand than stocking a few loose spares.
3. Many probe models and mixed thread types make it easy to order the wrong stylus
Most factories have more than one CMM, and the probe models vary widely. One rule to remember: the thread follows the probe, not the machine brand. Some domestic and imported machines use the same probe head and controller system, so saying “we use brand X” does not tell you the thread size.
The correct sequence is: identify the probe/probe head model first, then determine the thread, then choose the stylus tip diameter and length.
If both M2 and M3 probes are used on site, it is easy to pick the wrong one when ordering individually — an M2 stylus will not screw onto an M3 probe head, or if it does, the thread engagement is incorrect, causing drifting measurements at best and a stylus crash at worst. When buying as a kit, the set is already sorted by thread: 03969075 is the M2 kit, 03969040 is the M3 kit, and M00-694-008-000 is the M3-specific kit for scanning probes such as the HP-S-X1C, which also includes a suction cup and a star base. With the kits stored in separate compartments in the cabinet and labelled, the chance of picking the wrong one is far lower than with loose styli.
4. Scanning probe configuration: buying as a kit avoids a lot of matching hassle
Stylus configuration for scanning probes is more complex than for trigger probes. In addition to straight styli and extension bars, it involves accessories such as suction cups, adaptors and star bases. For example, M00-694-008-000 — according to the official documentation, this kit includes a suction cup and a star base, and is intended for scanning probes such as the HP-S-X1C.
When buying scanning styli individually, you have to work out for yourself: whether the suction cup model is correct, whether the adaptor is included, how many directions the star base needs to cover, and how much overhang the extension bar will give. Every additional joint adds a potential bending point and a potential loosening point. This is the selection principle the manufacturer repeatedly stresses — use the shortest stylus that meets the application and minimise the number of joints.
When you buy a kit, the manufacturer has already worked out the combinations for you, so compatibility within the kit is guaranteed. That said, the actual measuring range, probe configuration and workpiece geometry on a given machine still depend on the machine’s real setup. A kit covers the common combinations, not every possible one.
5. Ledger and stocktaking: kits are easier to manage than loose styli
For asset management, loose styli are a problem because each one has low value, there are many of them, and they are easy to lose. A stylus can sit in a tool cabinet for three months without being touched, and at stocktaking nobody can tell whether it is a spare or scrap. When you buy a kit, the whole set carries one barcode or one serial number, and storage, issue and stocktaking are all done per kit, so the records stay clear.
A more practical benefit is that missing parts are easy to spot. If an extension bar is missing from a kit, you see it the moment you open the box. With loose styli, you may not notice until you actually need one. This is covered in more detail in Section 6, “Maintenance of the kit”.
6. Switching styli across systems: the cross-reference table is a ready-made bridge
If the shop floor has both A-xxxx-xxxx and 0396xxxx part numbers, looking up numbers when mixing them is tedious. Page 34 of the official catalogue has a Renishaw ↔ Hexagon cross-reference table. We checked all 13 sample groups line by line, and it can be used directly. For example, A-5000-3551 corresponds to 03969301, described as M3-Stylus (R-1-SS-0.7-L21).
Note, however: combinations not listed in the table cannot be assumed to correspond. The comparison table is the official bridge, not a universal translator. When buying a kit, the part numbering system inside the kit is consistent, which removes the need to cross-reference between systems.
Summary of the decision criteria: unstable operating conditions, tight spare parts turnover, mixed probe models, complex scanning configurations, and strict inventory requirements — in these five situations, buying a kit is usually more convenient. Conversely, if the application is single, usage is predictable, and a dedicated person manages spare parts, buying individual items is more precise. The next section covers “when individual purchase is mandatory”.
Buying a kit is convenient, but only if you can use everything in it. When the working conditions do not match the kit’s default configuration, the kit becomes idle stock. The following situations require individual purchase.
1. Probe thread does not match the kit
Stylus kits are grouped by thread: M2, M3, M4 and M5 each form their own system. The thread follows the probe, not the machine brand. Identify the probe/probe head model first, then determine the thread: TP2, TP20, TP200 and REVO use M2; TP1, TP6, SP25M and REVO use M3; TP7M and machine-tool probing use M4; high-accuracy scanning uses M5. This correspondence is specified in the official catalogue. The M4 kit approach generally does not work — the Hexagon catalogue lists only one page of M4-thread styli and accessories, while Renishaw’s M4 range targets TP7M and on-machine probing, so ordering individual styli as needed is the norm.
2. Special geometry of the feature being measured
Kits are mainly made up of straight styli and extension bars, covering conventional holes, faces and steps. Deep holes, narrow grooves, internal threads, curved sidewalls and disc-shaped end faces require special components such as stepped bars, disc styli, star styli and articulating joints, e.g. M2-Disc Stylus (R-6-SS-L10), M2-Star Stylus 5-way (R-2-SS-1.4-SP20/30) and M2-Stylus - Stepped (R-0.5-TC-0.32/0.7-L10), which are usually not in the default kit list. List the features on the drawing and order the styli for each special geometry individually.
3. Stylus tip material may need to be changed
Ruby is suitable for the vast majority of applications. Heavy-load scanning of aluminium parts carries a risk of adhesive wear; the manufacturer recommends silicon nitride. Heavy-load scanning of cast iron carries a risk of abrasive wear; the manufacturer recommends zirconia. These two stylus tip materials are not included in the standard kits. For triggered point measurement with short contact time, ruby is generally sufficient. For continuous scanning of aluminium or cast iron workpieces, follow the two recommendations above. The material codes in the specification table are clearly marked (R = ruby, TC = tungsten carbide stem, SS = stainless steel stem, CE = ceramic, etc.); check the codes when ordering.
4. Hard constraints on length and stem diameter
Extension bars in the kits come in fixed lengths, e.g. the M2 kit 1 includes extension bars of 5/10/20/30 mm. If the workpiece depth falls between two standard lengths, or the required reach exceeds the kit’s assembly range, you must order a separate extension bar. The principle is: use the shortest stylus that meets the application and minimise the number of joints — every connection between stylus and extension bar adds a potential bending point. EWL is measured from the stylus tip centre to the point where the stylus stem interferes with the feature being measured; total length L is measured from the mounting face to the stylus tip centre. These are listed separately in the specification table, so check whether the EWL is sufficient when selecting. If the required EWL falls outside the kit’s coverage, a single straight stylus of the appropriate length is often more stable than assembling an extension bar.
5. Keep separate records for spare parts and wear parts
Styli are consumable items and must be reordered individually. From the outset, register each stylus in the kit separately: part number, material, ball diameter, shank diameter, L, and EWL — one line per stylus, aligned with the specification table fields. When a stylus is damaged, simply reorder it by part number. The cross-reference table is the official bridge — for example, A-5000-3551 ↔ 03969301 and A-5000-7806 ↔ 03969201 — and can be used directly. However, combinations not listed in the table must not be assumed. For cross-system replenishment, first consult the table; if the part is not listed, find a stylus of the same specification by description. The description format is Thread - Type (Ball Material - Ball Diameter - Shank Material - Shank Diameter - L Overall Length), e.g. M3-Stylus (R-3-SS-1.5-L21) means an M3 straight stylus, ruby ball Ø3, stainless steel shank Ø1.5, overall length 21 mm. This naming convention can be used to verify specifications, but must not be used to derive part numbers in reverse.
6. Dedicated components for scanning probes
Stylus kits for scanning probes (e.g. HP-S-X1C) include a magnetic plate and star stylus holder, which differ in structure from trigger probes. In the scanning probe catalogue, the M3 section lists “magnetic plate and adaptor” and “stylus kit” separately, and the M5 section does the same. If the magnetic plate specification or adaptor type does not match the kit default, the components must be ordered separately. Note: it is not true that “scanning probes do not use threads” — M3 and M5 both have threaded versions in the scanning range; D2 is the non-threaded type. Refer to the probe model and the catalogue section to determine the correct type.
7. Boundaries for mixed kits and replacements
Mixing different kits, or using old parts with new kits, is acceptable as long as the threads match and the specifications align. However, combinations not listed in the comparison table must not be assumed to correspond. The two part-number systems are inherently different (Renishaw A-xxxx-xxxx, Hexagon 0396xxxx and M00-xxx-xxx-xxx), so cross-system replacements must follow the official comparison table. If a combination is not listed, verify the material, stylus tip diameter, stylus diameter, and length item by item against the description; if they cannot be confirmed, do not substitute.
Summary of the selection process
| Step | Check item | Outcome |
|---|---|---|
| 1 | Probe/probe head model → thread | If the thread does not match the kit, order a single stylus |
| 2 | List of workpiece features | Special geometry requires a dedicated stylus |
| 3 | Workpiece material and scanning load | Heavy-duty scanning of aluminium/cast iron: change stylus tip material, order single stylus |
| 4 | Calculate EWL and required overhang | If the kit length is insufficient, order a single straight stylus |
| 5 | Spare parts ledger | All future additions are ordered as single styluses and logged by part number |
Refer to the official data sheet for exact stylus tip diameter, stylus diameter, and length values. When selecting, cross-check the workpiece drawing, probe model, and catalogue specification table together.
Buying a stylus kit is only the start. What really affects how efficiently it works is its condition after being broken up, mixed with other kits, or “borrowed for a moment” on the shop floor. This section covers three things: how to replace missing parts, how to prevent mixing, and how to keep a stock record.
The items most likely to go missing from a stylus kit are not the styli themselves, but the extension bars and small accessories. Before ordering a replacement, first determine which thread system the missing item belongs to. The sequence is: identify the probe head/probe head model → then the thread → then the ball diameter and length. The thread follows the probe head, not the machine brand. A common mistake on site is ordering by “our machine is brand X”, which results in the wrong thread.
Once the thread is confirmed, check the specification against the official catalogue fields: part number / material / L (total length) / D (ball diameter) / S (shank diameter) / EWL (effective working length) / W (weight). For a straight stylus, at least match ball diameter D, shank diameter S, and total length L. For an extension bar, the key points are length and the thread at both ends. Extension bars are listed separately in the catalogue: M2 on page 19, M3 on page 28, scanning-type M3 on page 42, and M5 on page 57. When looking them up, do not compare against the stylus pages.
If the original packing list or label is still available, the easiest way is to order replacements by part number. If not, you can work out the specification from the official naming convention, e.g. M2-Stylus (R-1-SS-0.7-L10) means an M2 threaded straight stylus, ruby ball Ø1 mm, stainless steel stem Ø0.7 mm, overall length 10 mm. The naming convention is only a general guide and does not mean every part number follows the same coding logic; always refer to the official data sheet.
If many stylus components are missing, you can order a complete stylus kit from the catalogue: the trigger-type M2 stylus kit 1 (part no. 03969075) includes straight styli with various ball diameters and extension bars in lengths 5/10/20/30; the M3 stylus kit 2 (part no. 03969040) includes M3 straight styli and extension bars; the scanning-type M3 stylus kit (part no. M00-694-008-000) includes suction cups and star adaptors. The exact number of styli in each kit is given in the official data sheet; the catalogue layout uses columns, so do not count items by eye.
Mixing styli will not immediately cause measurement errors, but it makes it impossible for anyone to say what a given stylus is for. This happens at two levels.
The first level is thread mixing. M2, M3, M4, and M5 threads are all different and correspond to different probe/probe head systems. The applicable wording in the official Renishaw brochure is: M2 for TP2, TP20, TP200, and REVO™; M3 for TP1, TP6, SP25M, and REVO™; M4 for TP7M and on-machine measurement; M5 for the highest-precision scanning probes. Putting M2 and M3 styli in the same box and relying on feel when changing them wastes time and risks damaging the threads. It is recommended to label the thread size directly on the storage box rather than just writing “stylus kit”.
The second level is mixed usage. Under the same thread, straight styli, disc styli, star styli, and stepped rod styli serve different purposes. In the comparison table, the M2 disc stylus (03969241, M2-Disc Stylus (R-6-SS-L10)) and the M2 star stylus (03969056, M2-Star Stylus 5-way (R-2-SS-1.4-SP20/30)) are completely different structures: the former is used to measure features such as grooves and edges, while the latter is used for multi-directional triggering. Organising by type and labelling by specification saves more effort than counting afterwards.
The suction cups and adaptors in scanning stylus kits are not interchangeable with trigger-style accessories. The “suction cup and adaptor” for scanning M3 is on page 44 of the catalogue, and for M5 on page 61; they are listed in different sections from the trigger-style “stylus accessories”, so it is also recommended to store them separately.
Part numbers from different brands cannot be mapped to each other. Renishaw’s A-xxxx-xxxx and Hexagon’s 0396xxxx / M00-xxx-xxx-xxx are two separate numbering systems; the official cross-reference table is the only bridge. Any combination not listed in that table must not be assumed. If you come across a part on site and cannot tell which brand it belongs to, set it aside rather than grouping it by appearance.
The ledger must answer three questions: what is in the kit now, what is missing, and who is using it. Create one row per kit:
| Field | Record content | Purpose |
|---|---|---|
| Kit / part number | e.g. 03969075, 03969040, M00-694-008-000 | Matches the catalogue; look up directly when ordering replacements |
| Thread system | M2 / M3 / M4 / M5 | Prevents mixing threads and ordering the wrong type |
| Compatible probe / probe head | Fill in according to the actual site configuration | Quickly confirms whether the kit works with a given probe |
| Contents list | List each part number or description item by item | Basis for stocktaking and replenishment |
| Status | In use / on loan / to be replenished / retired | Keeps the movement of kits traceable |
For the “existing parts list”, it is recommended to use the official naming convention, e.g. M3-Stylus (R-3-SS-1.5-L21). Even if the original label is lost, the part can still be identified against the catalogue by its description. The cross-reference table provides the official mapping, e.g. A-5000-3553 ↔ 03969303, described as M3-Stylus (R-3-SS-1.5-L21), which can be cited directly.
Updating the ledger is tied to two actions: log every loan/return, and do a full count after each measurement task. Real-time updates are not required, but any “missing parts” must be caught before the next use—not discovered on the shop floor when an extension bar is found to be missing.
For scanning stylus kits, the condition of the suction cup and adaptor must be recorded separately in the ledger. These components are frequently removed and refitted when changing probes, so they are more prone to wear and loss than straight styli. Listing them on a separate line makes them easier to manage than lumping them under “accessories”.
For any ledger fields involving specific dimensions, quantities or weights that are not explicitly listed in the catalogue, do not fill in figures from memory. Mark them as “subject to official data sheet” and complete them once the official data is available. The value of a ledger lies in accuracy, not completeness—entering the wrong stylus tip diameter is more troublesome than leaving the field blank.
First identify the probe, then determine the thread: TP20 belongs to the M2 system, corresponding to item 03969075 (M2 stylus kit 1) in the Hexagon catalogue. This kit includes straight styli with various ball diameters and extension bars in lengths of 5/10/20/30 mm, covering common trigger measurement applications. If your probe is TP1, TP6, SP25M, REVO, or another M3 system, you need to switch to 03969040 (M3 stylus kit 2). Note the sequence: probe → thread → ball diameter and length. Do not infer the thread directly from the machine brand—many domestic CMMs and some Mitutoyo models use Renishaw-style probe heads. For the exact number of styli and specifications in each kit, refer to the official data sheet.
Not necessarily — it depends on the model of scanning probe you are using. M00-694-008-000 is the M3 scanning-specific stylus kit listed in the catalogue as “suitable for HP-S-X1C”. It includes a suction adaptor and a star-shaped holder, and is intended for scanning applications that require multi-directional probing. If your scanning module is an M3 system such as SP25M or REVO, but you are not using HP-S-X1C, the standard 03969040 (M3 stylus kit 2) will cover your needs for straight styli and extension bars. Scanning probes in the M5 system (e.g. the highest-accuracy scanning type) have their own dedicated M5 stylus section, which is not covered by the three kits mentioned above; these must be selected individually using the specification tables on pages 51–64 of the catalogue. In short, check the probe head model before ordering a kit — do not order based on the word “scanning” alone.
According to the catalogue description, a typical KIT consists of straight styli in various ball diameters plus extension bars of different lengths. The scanning-specific kits (e.g. M00-694-008-000) also include a suction cup and a star stylus holder. The straight styli cover routine touch-trigger probing across different bore diameters and depths; the extension bars provide the reach needed for deep holes or obstacle clearance; and the star stylus holder enables multi-directional probing in a single setup. As for whether you will actually use all of them, it depends on your part family: if your workpieces vary widely in size and hole depth, buying the full kit saves the hassle of selecting individual components each time; if you only probe one specification over the long term, buying individual styli is more cost-effective. The catalogue is laid out in columns, so refer to the official data sheet for the exact number of styli in each kit rather than estimating from memory.
Buying individually is perfectly fine and more accurate. Taking the actual part numbers in the comparison table as an example, the M2 system Ø1 mm ruby ball, stainless steel stem Ø0.7, 10 mm overall length straight stylus is 03969201 (description: M2-Stylus (R-1-SS-0.7-L10)) — just order with this part number. The advantage of buying a kit (e.g. 03969075) is that it covers multiple ball diameters and extension bar lengths in one go, which suits new machine installations or applications where the workpiece family varies widely. If you already have a basic configuration and only need to fill a gap, buying a single stylus avoids waste. The deciding factor is simple: do you need “one stylus” or “a full set of capability”? If the former, buy individually; if the latter, buy the kit.
The comparison table is the official cross-reference, butonly combinations verified line by line in the table can be used; combinations not listed cannot be inferred. The trigger-type comparison table on page 34 of the catalogue (Renishaw ↔ Hexagon) has been verified line by line for 13 sample pairs, e.g. A-5000-7806 ↔ 03969201 and A-5000-3551 ↔ 03969301, which can be referenced directly. Note two points: first, the scanning-type comparison table on page 67 isITP ↔ Hexagonin two columns, with part numbers such asTH M3 050 04 052 ↔ M00-694-221-000, whichdo not include Renishaw numbers—do not mistake the ITP number for a Renishaw number. Second, for combinations not listed in the table, do not extrapolate from naming patterns; rely on the official data sheet.
Not recommended. The thread type follows the probe, and trigger and scanning probes are listed as two separate sections in the catalogue, each with its own numbering: trigger probes are covered in Chapters M2/M3/M4 (pages 14–32), while scanning probes are covered in Chapters D2/M3/M5 (pages 38–64). Even the suction cups and adaptors are exclusive to the scanning section (pages 44 and 61). Even for the same M3 thread, the scanning-specific kit (e.g. M00-694-008-000) includes a suction cup and a star stylus holder, whereas the trigger Kit 2 (03969040) does not. In practice, kits should be ordered separately according to the probe head model: keep 03969075 or 03969040 for trigger stations, and order the scanning kit separately to match the scanning module. Mixing the two adds extra joint faces, increases overhang, and introduces more potential loosening points — the drawbacks outweigh any benefit.
Turn the multiple-choice question into a process question, and the answer becomes obvious. When you start work tomorrow, you can follow the sequence below without having to re-read the whole document.
First, ask yourself: in the next twelve months, will you need to measure more than three types of features? If the answer is no, and each feature is measured frequently, then buying individual styli is usually the better choice — it keeps cost where it matters and avoids unused items. If the answer is yes, or you are unsure, then the marginal cost advantage of a kit starts to show, and you can move on to the second step.
Second, look at how often styli are changed on the shop floor. Every change requires recalibration and re-alignment, and these hidden time costs often far exceed the price of the stylus itself. If one workpiece on your production line requires three different feature types, the value of a stylus kit combined with an automatic exchange mechanism is not about “saving money” but “saving time”. Conversely, if a shift only requires one or two changes, the cost of manual stylus exchange is negligible, and the automation advantage of the kit is not realised.
Third, weigh budget and risk together. The risk of buying individual styli is that if requirements change, you have to source replacements at short notice, and the cost of downtime while waiting for stock can far exceed the initial saving. The risk of a kit is that it always contains a few styli you may use only once a year, tying up budget as well as storage and calibration effort. Which way the balance tips depends on how certain your requirements are — the more uncertain they are, the more the kit’s “insurance” value matters; the more defined they are, the more cost-effective the “precision” of individual styli becomes.
Finally, a quick self-check: write down every stylus you intend to buy, and next to each one note the specific workpiece and feature it is for. If you cannot write that down, you do not need it. This ten-minute exercise will prevent most impulse purchases. The value of a tool is never in its quantity, but in the moment the probe touches the workpiece — it is exactly the stylus you need.
Every machine tool has different configuration, operating conditions and budget; this article only provides a reference framework. You can:
Ningbo JCETech – 15 years focused on precision measurement, supporting the selection, installation, calibration, repair and maintenance of the full Renishaw, Marposs and Blum-Novotest range of probes and tool setters.
| Contact Information | Details |
|---|---|
| Official booking | jcetech.cn/contact |
| Service hotline | 157-5780-7400 / 132-2194-1413 |
| Repair shipping address | A2-12B17, Ningbo Yinzhou R&D Park |
| Service coverage | Nationwide on-site support · 48-hour response for urgent repairs |
Send us your problem and let our engineers work out the numbers for you.
If the criteria or procedures in this article do not match your machine, send us the control system, probe model and a photo of the installation — we will answer with the specific settings rather than a generic recommendation.
| Item | Detail |
|---|---|
| Company | NINGBO JCE INDUSTRIAL TECHNOLOGY CO., LTD. (JCETech) |
| Address | Room 14-1-17, Building 033, Building 2, No. 15, Lane 587 Juxian Road, Juxian Subdistrict, Hi-Tech Zone, Ningbo 315048, Zhejiang, P.R. China |
| Contact | Mr. Li / Mr. Zhang |
| Phone | +86 157-5780-7400 (Mr. Li), +86 132-2194-1413 (Mr. Zhang) |
| Online booking | jcetech.cn/contact |
Repair intake and on-site service use the same address. Please call before shipping a unit so transport and scheduling do not have to be repeated.
Probe, tool setter and laser interferometer repair, calibration and on-machine measurement integration.
157-5780-7400 / 132-2194-1413
打开微信扫一扫,关注「宁波匠测」
获取维修进度查询与技术支持
打开企业微信或微信扫一扫