Your CMM Actually Uses Renishaw Stylus Specs

Your CMM Actually Uses Renishaw Stylus Specs

Identify the probe head before the machine brand, use a three-step confirmation method, and match the thread to the probe model—M2, M3, M4, or M5—to avoid costly mistakes.

Introduction

“Master, is the stylus on our machine M2 or M3? I ordered the same model as the previous one, but when it arrived, it wouldn’t screw on.”

Almost every experienced CMM operator has said this at some point. Both machines are domestically made, yet the previous one took an M3 stylus while this one won’t tighten at all. Both are imported machines from the same brand, but after changing the probe head, the thread size is different. Worse still, you take the old stylus to match a replacement, and the ball diameter and stem diameter are right, but the thread is one pitch off — it fits loosely, and nobody trusts the measurement data.

Where does the problem lie? It comes from a common misconception: assuming the thread follows the machine brand. In fact, the stylus thread on a CMM is never determined by “which brand the machine is” but by the probe head and probe model. Many domestic CMM brands — and even some imported models — use Renishaw-style probe heads and probe modules internally. In other words, the “domestic machine” you have may well need Renishaw-spec styli.

That brings us to the core question this article addresses: when faced with a CMM, how do you quickly and accurately determine which thread size stylus it needs?

We’ll do this in three steps. First, identify the probe head before the machine itself — find out which machines actually carry Renishaw probe head platforms. Second, use the “three-step confirmation method” — check the nameplate, match the probe model, and confirm the thread specification — to lock down the thread on site within five minutes. Third, look up the official thread catalog by probe model, so M2, M3, M4, and M5 each fall into place — no guessing, no relying on experience, no “the one we had before”.

After reading this article, you’ll have a thread identification process that works on any brand of machine, without being misled by the machine nameplate. You’ll also understand the official probe-to-thread logic from M2 to M5 — knowing that TP20 uses M2, SP25M uses M3, and TP7M uses M4, rather than memorizing a list of part numbers. You’ll also get a practical reference for changing styli on mixed domestic and imported machines, plus the three most common thread selection mistakes — going by brand, by machine body, or by interface — so you can avoid the traps that trip up even experienced technicians.

Choosing the wrong thread for a stylus is not just a matter of it not screwing on. If the thread is wrong, forcing it can damage the threaded hole in the probe head. Even if it does go on, the stylus axis will be tilted, calibration data will drift, and every subsequent measurement will be built on a faulty foundation. Rather than troubleshooting after a problem appears, it’s better to confirm the specification before changing the stylus.

Let’s start, then, with “identify the probe head before the machine”.

1. First identify the probe head, then the machine: which machine carries the Renishaw platform

When a field engineer picks up a stylus, the first instinct is often to check the machine nameplate: a CMM of brand XX, then look up the stylus by brand. This habit is especially likely to lead nowhere on domestic machines. The reason is simple: many domestic CMM brands, as well as some imported models, use the Renishaw platform for the probe head, probe, and controller. Who made the machine casing and whose probe is installed inside are two different things.

The only reliable way to determine the thread size is by the probe/probe head model, not the machine brand. If you are working on a domestic machine and open the probe to find a TP20 module, it uses M2 thread styli; if it has an SP25M scanning module, it uses M3 thread. Conversely, an imported machine fitted with an older TP6 manual probe also uses M3 thread. The thread follows the probe, not the brand.

In Renishaw’s official thread catalogue, the compatible probes are clearly specified. The M2 thread section covers the TP2, TP20, TP200 and REVO™; the M3 section covers the TP1, TP6, SP25M and REVO™; the M4 section covers the high-accuracy TP7M and the machine-tool probe series; and the M5 section is for the highest-accuracy scanning probes. Note that REVO™ appears in both the M2 and M3 sections — this means that, within the same probe head system, different modules can fall under different thread specifications. So identifying by brand alone is not reliable; you must identify the specific probe model.

How do you confirm this quickly on site? Three steps. First, open the probe cover and look for a model marking on the probe head or probe. Models such as TP20, TP200 and SP25M are usually printed directly on the module or head. Second, check the thread diameter of the existing stylus. Take a known stylus, compare it with a thread gauge, or check the official catalogue to confirm the thread specification. This quickly rules out obvious mismatches. Third, and most reliably, check the machine documentation or contact the machine supplier to confirm the exact probe head model, then look up that model in the official thread catalogue.

Here is an important point to note: mixed configurations on domestic machines. Some domestically produced CMMs are built in-house, but the entire probe head, probe, and stylus system is purchased from Renishaw. In this case, if you look up the domestic parts catalog by the machine brand, you may not find the corresponding thread information, or the specifications listed may not match what is actually installed. The correct approach is to ignore the machine nameplate and check the probe head model directly. Similarly, some imported machines—for example, certain Mitutoyo equipment—also use Renishaw probe heads, so their stylus specifications follow the Renishaw system, not Mitutoyo’s own accessory system.

Here is another specific scenario. An equipment engineer is tasked with replacing the stylus on a domestically produced gantry CMM. The machine nameplate shows a domestic manufacturer, but when you open the probe compartment, you see a TP200 probe module. In this case, just consult the Renishaw M2 catalog; the machine brand is irrelevant. If you search for accessories by the domestic machine brand, you will likely find a different specification system, and the parts may not fit or may fail calibration after installation.

Another common point of confusion is Hexagon CMMs. Hexagon publishes its own stylus catalogue, split into two sections: “trigger” and “scanning”. The trigger section covers M2, M3 and M4 threads; the scanning section covers D2 (unthreaded), M3 and M5 threads. However, the Hexagon catalogue includes a dedicated cross-reference table (page 34) that maps Renishaw part numbers to Hexagon part numbers one-to-one. What does this tell us? Many Hexagon machines use Renishaw-standard stylus interfaces. Take a Renishaw A-5000-3551 (M3 threaded straight stylus) – you will find the corresponding Hexagon part number 03969301 in their catalogue. This is not a coincidence; it is an industry-standard interface. So even on a Hexagon machine, you should first confirm the probe model, then decide which catalogue to consult for thread specifications.

Another common misconception on the shop floor is assuming that an imported machine brand automatically means a particular stylus thread. For example, some engineers believe “Zeiss machines always use M3” – that is not accurate. Zeiss equipment has its own dedicated stylus catalogue, but the thread used on a given machine still depends on the probe installed. Probes of the same brand but different models can have completely different threads. The only reliable rule is to go by the actual machine configuration.

In practice, it is a good habit to record the probe head model and current stylus thread size on the equipment checklist before every stylus change. That way, the next time you swap styluses, you can simply check the record without having to remove the protective cover again. This is especially important in workshops with multiple machines—different machines may use different thread sizes, and mixing them up can easily lead to incorrect installation.

To summarise the decision logic of this section: first, ignore the machine brand and open the probe compartment to check the probe head model; second, look up the official thread catalogue for that probe head model to determine the thread size; third, select the stylus based on the thread size. Do not reverse this order. Identify the probe head first, then the machine—get the order right, and stylus selection will stay on track.

2. Three-step confirmation method: check the nameplate, verify the probe model, and determine the thread size

On site, the biggest concern when changing styluses is not choosing the wrong ball diameter, but the wrong thread. If the thread is wrong, the stylus simply will not screw on, and forcing it can damage the probe head interface. A more subtle issue is when the thread type is correct but the size does not match—for example, fitting an M3 stylus onto an M2 probe head may look similar, but they are completely incompatible. To avoid these problems, you do not need to memorise any manual—just follow the three steps below.

Step 1: Check the nameplate and confirm the probe head and probe model

Every CMM has a nameplate on its column or base showing the machine model and serial number. However, the machine model only tells you which manufacturer built the machine; it does not indicate the thread specification. What you actually need to record is the probe head and probe model—usually printed on the side or top of the probe head body as a combination of letters and numbers, such as TP20, TP200, or SP25M.

A common misconception is that buying a machine from a certain brand means the probes must also follow that brand’s specifications. In reality, many domestic CMM brands and some imported models use the Renishaw system for their probe heads and probes, so there is no fixed relationship between the machine brand and the thread specification. The only reliable way to determine the thread type is by the probe model, not the machine nameplate. Therefore, the key takeaway from this step is to note the probe model, not just the machine model.

Step 2: Match the probe model to the correct thread section in the catalog

Once you have the probe model, refer to the applicable instructions in the official thread catalog to determine which thread type to use. Common correspondences are listed below for quick reference:

Probe model (common) Thread specification Official catalogue description
TP2, TP20, TP200, REVO™ M2 For industry-standard CMM probes
TP1, TP6, SP25M, REVO™ M3 For manual and scanning probes
TP7M and machine-tool probing probes M4 Rigidity and weight optimised, designed specifically for high-precision probes
Highest-precision scanning probes (Renishaw and other manufacturers) M5 High-end range, designed specifically for the highest-precision scanning probes

Note that REVO™ appears in both the M2 and M3 rows. This is not a typo; it indicates that the same probe model can have different thread sizes, depending on the module actually fitted to the probe. Similarly, the SP25M scanning module uses M3, but if the same machine is fitted with a trigger module instead, the thread may become M2. The key action in step two is therefore: look up the catalogue by probe model, not by machine brand.

In addition, the Hexagon catalogue also separates trigger and scanning types by chapter: trigger probes use M2, M3 and M4 threads, while scanning probes use D2 (unthreaded), M3 and M5 threads. D2 unthreaded styli are not screwed in; they are located by a suction cup or adaptor. Pay particular attention to this when selecting a stylus, so that an unthreaded stylus is not mistaken for one with a damaged thread.

Step 3: Determine the thread specification, then look up the specific part number

Once the thread specification is fixed, the stylus selection range narrows considerably. The next step is not to look up part numbers directly, but to confirm three parameters first: stylus tip diameter, stem diameter, and overall length. These three parameters are determined by the feature being measured—a small tip diameter for small holes, a longer overall length for deep cavities, and a dedicated stylus for thread profiles. The official selection guideline is to use the shortest stylus that meets the application, minimise the number of joints, and use the largest tip possible. These are not empty words—they directly affect measurement accuracy.

For example, suppose you need to measure a blind hole with a diameter of 5 mm and a depth of 15 mm on the shop floor. If the probe head is a TP20 and the thread specification is M2, then, following the “largest tip possible” principle, the tip diameter, stem diameter, and overall length should be determined based on the feature size and the official selection guideline; refer to the official data sheet for the exact values. If the bottom of the blind hole also has a relief groove, a disc stylus is required—the thread is still M2, but the geometry is completely different.

Here is a special reminder: the same thread specification does not mean the styli are interchangeable. A straight stylus with M2 thread and a disc stylus with M2 thread share the same mounting interface, but their applications and calibration methods are completely different. A disc stylus typically only needs to be calibrated for one diameter and can only effectively probe in the X and Y directions; to probe in the Z direction, a stylus with a radius-end roller is required. These differences should be considered carefully during selection — do not place an order simply because the threads match.

Another common issue is the extension bar. The thread of the extension bar must match the stylus — an M2 stylus uses an M2 extension bar, and an M3 stylus uses an M3 extension bar. Each additional joint introduces another potential point of bending and deviation, so use an extension bar only when necessary. If extra reach is truly required, prefer a one-piece long stylus over a “short stylus + extension bar” combination.

The last step is to look up the part number. Note that Renishaw and Hexagon use different part-number systems: Renishaw part numbers look like A-5000-xxxx, while Hexagon part numbers come in formats such as 0396xxxx and M00-xxx-xxx-xxx. Always refer to the official data sheet for the exact part number. Both manufacturers’ official catalogs include cross-reference tables—for example, page 34 of the Hexagon catalog lists the Renishaw-to-Hexagon trigger-style stylus cross-reference. Combinations shown in the table can be matched directly; combinations not listed cannot be assumed to correspond. Always check the official data sheet rather than guessing from experience.

Once these three steps are complete—thread size, stylus construction, and ball diameter, rod diameter, and length all confirmed—ordering will be error-free. When changing the stylus on site, if the probe model and thread size are correct but the stylus feels unusually tight or loose when screwed in, first check the thread for burrs or debris, then re-verify with a thread gauge. Do not force it—once the thread strips, the probe head interface is ruined.

3. How to choose between M2 and M5: Check the official thread catalog by probe model

When determining thread size on site, the worst approach is to guess based on the machine brand. On the same CMM, changing the probe head module can switch the thread from M2 to M3. The thread follows the probe, not the machine brand—this is the first rule of selection.

In the official Renishaw catalogue, styli are grouped by thread into separate volumes for M2, M3, M4, M5, and Faro articulated-arm styli. Each volume begins by listing the probe models it covers. This cross-reference is the anchor point for selection—remember it and you will avoid detours on the shop floor.

M2 Volume covers the TP2, TP20, TP200, and REVO™ probes. All three TP-series probes are trigger probes and are very common on domestic machines and older imported models. If you see a TP20 module or a TP200 probe body on the probe head, the thread is almost certainly M2. When the Equator comparator is fitted with a TM25-20 module (the TP20 trigger module), it also uses M2 styli.

M3 Volume covers the TP1, TP6, and the SP25M scanning probe, as well as REVO™. TP1 and TP6 are classic trigger probes from the era of manual indexing probe heads and are still widely used today. The SP25M is a scanning probe; when fitted with the SM25-2 scanning module, it uses M3 styli. One point is easy to confuse: REVO™ appears in both the M2 and M3 volumes because it can accept both trigger and scanning modules, and the thread depends on which module is fitted. So when you see a REVO™, you cannot determine the thread directly—you need to open it up and check which module is currently installed.

M4 Volume covers the TP7M and Renishaw’s machine-tool probing probes. The TP7M is a high-accuracy trigger probe, and the M4 thread was designed with rigidity and weight optimisation in mind. The OMP40-2, OMP400, OMP60, and OMP600 commonly found on machining centres, the LP2 on lathes, and the TS27R tool setter all use M4 threads. If you are selecting styli for machine-tool probes in the workshop, go straight to the M4 volume.

M5 VolumeThe probe covered in this volume is the highest-precision scanning probe from Renishaw and other manufacturers. The M5 booklet is intended for the highest-precision scanning probes; specific stylus diameter and rigidity parameters are as per the official data sheet. In the Hexagon catalogue, M5 threaded styli and accessories span pages 51 to 64, with considerably more content than the M4 volume, which is consistent with the M5 volume’s focus on scanning applications.

Make a quick-reference table of the correspondence between these volumes and post it next to the workstation:

Thread specification Corresponding probe model (official brochure description) Typical application scenarios
M2 TP2, TP20, TP200, REVO™ (trigger module) General-purpose trigger measurement, Equator comparator
M3 TP1, TP6, SP25M, REVO™ (scanning module) Manual probe head triggering, general-purpose scanning
M4 TP7M, machine-tool probing probe series High-precision triggering, machine tool alignment, and workpiece measurement
M5 Highest-precision scanning probes from Renishaw and other manufacturers High-precision scanning, complex surface measurement
Faro arm Faro portable articulated-arm CMM Styli specific to articulated arms

The actual identification sequence is: first identify the probe/probe head model, then determine the thread, and finally the stylus tip diameter and length. For example, on a machine on site, if the probe head is marked TP200, the thread is M2; if it is marked SP25M, the thread is M3. If the nameplate is illegible, remove the stylus and compare it with a thread gauge, or check the official catalog to confirm the thread specification.

Another common pitfall: the Hexagon catalog divides styli into two categories: “trigger” and “scanning”. The trigger category includes M2, M3, and M4 threads, while the scanning category includes D2 (no thread), M3, and M5. The D2 stylus is not screwed in by thread; it is located using a suction cup and adaptor. Do not confuse it with M2. Although the M3 in the scanning category and the M3 in the trigger category share the same thread specification, the stylus stem diameter, length, and tip diameter ranges differ. When consulting the catalog, look in the corresponding section for each type.

Finally, one key point: thread size is only the first step. Both the TP6 trigger probe and the SP25M scanning probe use M3 threads, but they differ in the stylus length and weight they can support. Scanning probes are more sensitive to stylus overhang, and longer styli amplify bending errors. For a given probe model, the maximum stylus length and combination weight are specified in the official data sheet. When selecting a stylus, keep in mind the three principles the manufacturer repeatedly stresses: use the shortest stylus that meets the application; minimise the number of joints (each additional joint adds a potential bending point); and use the largest stylus tip diameter possible (this increases the clearance between the tip and the stylus stem, reducing false triggers caused by the stem contacting the workpiece). These three principles will help you avoid most on-site problems more effectively than any comparison chart.

4. Mixing domestic and imported machines: a comparison table to check before changing styli on site

As the number of machines on site grows, it is common to have both domestic and imported CMMs in the same workshop. Many engineers get into the habit of taking a stylus from one machine and fitting it to another when one runs short. This works fine when the thread sizes match, but once you start mixing machines, problems can easily arise.

The problem stems from a deep-rooted misconception: many people assume that thread specifications follow the machine brand—domestic machines use domestic styli, and imported machines use imported styli. In fact, the thread specification follows the probe model, not the machine brand. Many domestic CMM brands use the Renishaw system for their probe heads and controllers, and their probe models correspond directly to Renishaw’s TP or SP series. Conversely, some imported machines may also feature probe heads that are not based on the Renishaw system. So, neither “this is a domestic machine” nor “this is an imported machine” can directly determine which thread type to use.

The correct order of determination is: first identify the probe/probe head model, then determine the thread specification, and finally decide on the stylus ball diameter and length. Before changing styli on site, even for a temporary swap, it is recommended to follow this sequence.

The procedure is carried out in three steps. First, check the probe head or probe body for a model label. Renishaw touch-trigger probes such as the TP2, TP20 and TP200, and scanning probes such as the SP25M, normally have the model printed directly on the probe body. If the label is worn and unreadable, consult the machine manual or contact the equipment manufacturer to confirm. Second, once the probe model is identified, refer to the thread catalogue to determine the thread size. The Renishaw catalogue states clearly: M2 thread is used for TP2, TP20, TP200 and REVO™; M3 thread is used for TP1, TP6, SP25M and REVO™; M4 thread is designed specifically for TP7M and on-machine inspection probes; M5 thread is used for the highest-accuracy scanning probes. Third, after confirming the thread size, select the stylus ball diameter and length according to the size of the feature being measured.

On site, M2 and M3 are the most easily confused. Visually, the M2 thread diameter is clearly smaller than M3, but unless the two styli are compared side by side, judging from a single stylus alone is often uncertain. A more reliable method is to use a thread gauge, or simply try a stylus of known specification — if it screws in easily, it is the same thread size; if it will not go in, do not force it, as this can damage the thread in the probe head.

One point worth highlighting is the cross-reference table in the Hexagon catalogue. On page 34 of Hexagon’s “Styli and Accessories” catalogue, there is a Renishaw ↔ Hexagon comparison table that lists the corresponding Renishaw and Hexagon part numbers line by line. This table is the official reference, so it is the most reliable source to consult before changing styli on site. Here are a few practical examples:

Renishaw part number Hexagon part number Description
A-5000-3551 03969301 M3-Stylus (R-1-SS-0.7-L21)
A-5000-3552 03969302 M3-Stylus (R-2-SS-1.4-L21)
A-5000-3553 03969303 M3-Stylus (R-3-SS-1.5-L21)
A-5000-7806 03969201 M2-Stylus (R-1-SS-0.7-L10)
A-5000-7807 03969202 M2-Stylus (R-2-SS-1.1-L10)
A-5000-3604 03969203 M2-Stylus (R-3-SS-1.5-L10)
A-5000-3603 03969212 M2-Stylus (R-2-SS-1.3-L20)
A-5000-7805 03969220 M2-Stylus – Stepped (R-0.5-TC-0.32/0.7-L10)
A-5000-7808 03969221 M2-Stylus (R-1-TC-0.7-L20)
A-5000-3611 03969241 M2-Disc Stylus (R-6-SS-L10)
A-5000-3626 03969056 M2-Star Stylus 5-way (R-2-SS-1.4-SP20/30)
A-5000-3709 03969426 M4-Stylus (R-6-CE-4.5-L50)
A-5000-3627 03969054 M2-Styli Holder 5-way (SS-L7.5)

The value of this table is that it gives the official cross-reference between the two part-numbering systems. Renishaw part numbers take the form A-xxxx-xxxx, while Hexagon part numbers use formats such as 0396xxxx and M00-xxx-xxx-xxx. The two numbering schemes share no pattern, so they cannot be guessed. Combinations listed in the table can be used directly; combinations not listed must not be assumed to correspond — keep this in mind.

Another common situation on site: the probe fitted on a domestic machine tool is from the Renishaw system, but the consumables list supplied by the machine builder uses domestic part numbers. In this case, do not rely on the part number alone — check the thread specification. For example, if the list says “M3 straight stylus R-3-SS-1.5-L21”, it corresponds to Renishaw A-5000-3553, which is Hexagon 03969303. All three part numbers refer to the same stylus, just under different numbering systems. Conversely, if the list gives only a part number without a thread specification, confirm the thread before issuing the material.

Mixing scanning probes requires extra care. The Hexagon catalogue has a scanning comparison table starting on page 67, but that table only covers ITP ↔ Hexagon and does not include Renishaw part numbers. The table of contents lists it as “Renishaw | ITP | Hexagon”, but the actual page has no Renishaw column. There are separate Renishaw ↔ Hexagon comparison tables for scanning probes; the verified cross-references include: A-5003-0069 ↔ M00-1121-020-000 (M3-Stylus R-5-CE-2.5-L50), A-5003-4686 ↔ M00-1125-002-000 (M3-Rotary Stylus Knuckle), A-5003-4862 ↔ M00-694-241-000 (M3-Stylus R-8-CF-5-L75). When looking up scanning probe cross-references, first confirm which table you are using — do not use the ITP table to match Renishaw part numbers.

Another common pitfall: M4 threads. The Renishaw M4 brochure explicitly states that M4 is “designed specifically for the high-precision TP7M and machine-tool probing systems”, while the Hexagon catalogue devotes only one page to M4. In both ranges, M4 is a small but dedicated specification. If you come across an M4 stylus on site, do not underestimate it just because it takes up little space in the catalogue, and do not mix M4 with M3. M4 is M4 — if the threads do not match, it will not fit.

Finally, here is a quick reference table for the shop floor. It fits well on the inside of the toolbox lid or next to the probe head:

On-site actions Mandatory checks If the information is unavailable
Borrowing a stylus from a domestic machine for use on an imported machine Confirm that the probe models on both machines are identical Go by the probe model, not the machine brand
Borrowing a stylus from an imported machine for use on a domestic machine Confirm the thread size (M2/M3/M4/M5) Use a thread gauge or try a stylus of known thread size
Ordering a new stylus by part number Confirm which system the part number belongs to Check the official cross-reference table; do not assume compatibility if it is not listed
Changing the stylus on a scanning probe Confirm whether the probe is a trigger or scanning type, and use the corresponding cross-reference table For scanning types, first check the ITP ↔ Hexagon table starting on page 67; the Renishaw ↔ Hexagon scanning cross-reference appears on later pages—do not mix them up

Mixing styluses is not a problem in itself—the problem is doing so without confirming compatibility first. The thread follows the probe, not the machine brand. Remember this and you will avoid a lot of trouble on the shop floor.

5. The three most common causes of choosing the wrong thread: relying on the brand, the machine body, or the interface

Case 1: Relying on the brand—assuming that because the machine is from a certain brand, the thread must be a certain size

This is the most common type of misjudgment. Many engineers have a domestic CMM with a domestic brand name on the nameplate, so they assume that “domestic machines must use M3”, or conversely, seeing an imported brand, they assume “it must use M2”. Neither inference holds up.

Thread specification does not follow the machine brand—it follows the probe model. Renishaw’s official thread guide is explicit: M2 threads are used on probes such as the TP2, TP20, TP200, and REVO; M3 threads are used on the TP1, TP6, and SP25M scanning probes; M4 threads are reserved for the TP7M and machine-tool probing systems; M5 threads are for high-accuracy scanning probes. On the same CMM, if the probe head or probe module has been changed, the thread can shift from M2 to M3, or even to M5. The machine nameplate does not record this—it only shows the machine model and date of manufacture.

A subtler case: many domestic CMMs use Renishaw probe heads and controllers internally, and some imported models (including certain Mitutoyo machines that are not Renishaw-branded) also carry Renishaw probes inside. So saying “I have brand XX” tells you nothing about the thread. The correct approach is to lift the dust cover and read the probe model printed on the head—if it is a TP20, it is M2; if it is an SP25M, it is M3. The brand on the machine housing is irrelevant.

Scenario 2: Judging by the machine body—assuming “same machine model, so the thread used last time will work this time”

Factories often have more than one CMM of the same model, and identical units may have been purchased in separate batches. Some operators note the stylus type used on one machine and simply apply the same stylus to another machine of the same model. This habit is usually fine when the machine is completely unmodified, but if one unit has had its probe head replaced, its probe module upgraded, or its probe adaptor changed during service, the thread size may differ.

For example, two machines may look identical. One left the factory with a touch-trigger probe and an M2 thread; the other was later fitted with a scanning probe module for scanning work and now has an M3 thread. From the outside, the columns, beams and worktables look the same, but the styli are not interchangeable at all. If you rely on the machine’s appearance, you may buy a stylus that will not screw in, or worse—the thread is close enough to engage a few turns, but the stylus works loose during measurement, causing data drift. It can take a long time to trace the problem back to a thread mismatch.

The rule is simple: do not remember “which stylus goes with which machine”; remember “which stylus goes with which probe module”. Keep a card for each machine listing the current probe model and its thread size. Before changing a stylus, check the card, then verify the model marking on the probe body itself. If the machine has been serviced, update the card at the same time—otherwise the card itself becomes outdated information.

Scenario 3: Recognizing the interface – assuming “if it screws in, it’s correct”

This is the most dangerous case. The four thread types M2, M3, M4, and M5 have increasing nominal diameters, so they should not be easily confused. However, during on-site operation, someone may hold an M3 stylus and try to screw it into an M2 threaded probe head. Since M3 is thicker than M2, it will not fit—this is not a serious issue, at worst it simply cannot be installed. The real problem lies with certain adaptors between M2 and M3, or probe heads with built-in thread adaptors that look identical but have different thread specifications.

Another common scenario involves stylus extension bars. One end of the extension bar connects to the probe head, and the other end connects to the stylus; the thread specifications at each end may differ. For example, an M2-to-M3 extension bar has an M2 end for the probe head and an M3 end for the stylus. If you only remember “this machine uses M2” and see that the other end accepts an M3 stylus, you might assume the entire bar is M3. The next time you try to mount it on an M3 probe head, the M2 end will not fit—or worse, the M2 end may barely thread into a worn M3 hole, damaging the threads.

How to check: do not judge thread size by “feel”. Whether the thread matches depends on the official catalogue for the probe model. The Renishaw catalogue is organised by thread size (M2, M3, M4, M5), and each chapter states the applicable probe models at the start. The Hexagon catalogue is also organised by thread size, with a cross-reference table to Renishaw part numbers on pages 34 and 67. On site, use a thread gauge or a stylus of known thread size to confirm the thread before fitting the stylus. If the stylus feels “tight but will go in” when screwed in, do not force it — forcing in a stylus with the wrong thread will damage the thread hole in the probe, and this damage is irreversible; the only remedy is to replace the probe head or the probe itself.

Common root cause of the three cases, and one guiding principle

The common root cause of the three misjudgements is treating the thread as a property of the machine or the brand, rather than of the probe. The thread is part of the probe interface: change the probe and the thread changes with it; keep the probe and the thread stays the same. Machine brand, body appearance and interface feel are all unreliable indicators of thread size. The only reliable reference is the probe model.

There is only one decision path to follow: first identify the probe model → then check the official catalog → determine the thread specification → then select the ball diameter and length. The probe model is printed on the side or top of the probe head, usually in small text such as TP20, TP200, SP25M, etc. Once you have the model, open the corresponding chapter in the Renishaw catalog or the Hexagon brochure, confirm the thread specification, and then choose the stylus ball diameter, stem diameter, and overall length based on the features to be measured. The whole process does not require knowing the machine tool brand or recalling which stylus was used on the previous machine.

If the probe model is worn and unreadable on site, you can try a stylus of known specification to test the thread — but there is a prerequisite: this “known specification” stylus must come from official packaging or a reliable source, not one grabbed at random from a parts box. It is common for M2, M3, and M4 styli to be mixed together in a parts box. Using a stylus of unknown origin to test the thread is like using one unknown to verify another.

One final point: M4 threads are a “small but specialized” size in both systems. Renishaw’s M4 brochure states that it is dedicated to the TP7M and on-machine inspection probes; in the Hexagon catalog, M4-thread styli and accessories together take up only one page. If you come across a machine probe or high-accuracy CMM on site whose thread looks thicker than M3 but thinner than M5, don’t guess—look up the official brochure for the probe model. M4 has a narrow range of applications, but using the wrong one is costly: forcing it in will damage the threads, and M4 probes typically have a longer service interval than M2/M3.

FAQ

Q1. Our CMM is domestically made, and the nameplate says it’s a domestic brand, but I can’t read the label on the probe head. Should I just buy styli from the domestic brand’s accessory catalog?

Answer: No. Do not buy based on the CMM brand. The stylus thread size is not directly related to the CMM brand; it depends only on the probe head or probe model fitted on the probe head. Many domestic CMM brands use Renishaw-style probe heads and probes internally, so the thread size is naturally Renishaw’s. If you look it up in the domestic CMM brand’s catalogue, you may not find it, or if you do, the correspondence may have been passed through several hands, which is error-prone. The correct approach is three steps: first, check the machine nameplate to confirm the manufacturer and model—this is just for reference when looking up information later. Second, and most importantly, locate the model label on the probe head or probe body, and identify whether it is TP20, TP200, SP25M, or another model—this model directly determines the thread size. Third, take that probe model and look it up in Renishaw’s official thread chart to see whether it corresponds to M2, M3, M4, or M5. Remember this: the thread follows the probe, not the CMM brand.

Q2. I have a machine with TP200 marked on the probe head, but I have a batch of M3-threaded styli on hand. Can I screw them on directly?

Answer: No, you cannot screw it on directly. The TP200 uses an M2 thread, not M3. The official Renishaw catalogue states clearly that the M2 thread range applies to the TP2, TP20, TP200 and REVO probes, while the M3 thread range is for the TP1, TP6 manual probes and the SP25M scanning probe. So the TP200 must use M2-thread styli. If you only have M3 styli on hand, you either need to get M2 styli or consider an adaptor—but an adaptor adds an extra mating joint, which introduces another potential bending point and loosening point, and this affects accuracy. It is not recommended unless there is no other option. In addition, M4 thread is for the TP7M and on-machine probes, and M5 is for the highest-accuracy scanning probes; neither of these has anything to do with the TP200. The only basis for determining the thread specification is the probe model—nothing else counts.

Q3. We have a Hexagon machine in our workshop. We have always bought Renishaw styli, but recently our purchasing department said Hexagon sells its own styli. Can we mix styli from these two brands?

Answer: Yes, they can be mixed, but only if you find the corresponding part number in the official cross-reference table—you cannot infer it on your own. On page 34 of Hexagon’s “Styli and Accessories” catalog, there is an official “Renishaw ↔ Hexagon” cross-reference table. For example, Renishaw’s A-5000-3551 corresponds to Hexagon’s 03969301; both are M3-thread straight styli with a ruby stylus tip Ø1mm, stainless steel rod Ø0.7mm, and a total length of 21mm. Renishaw’s A-5000-7806 corresponds to Hexagon’s 03969201, an M2-thread straight stylus. Note that this table covers trigger probes only; there is a separate table for scanning probes, starting after page 67. That section includes an ITP-to-Hexagon comparison, and a few pages later a Renishaw-to-Hexagon scanning probe table. So first determine whether your probe is a trigger or scanning type, then consult the corresponding table. For combinations not listed, do not assume an equivalent—select based on each manufacturer’s official data sheet.

Q4. The workpieces I measure are often quite deep, so I want to buy longer styli. Is longer always better? Are there any limitations?

Answer: Longer is not always better—in fact, the first rule in Renishaw’s official selection guidance is to use the shortest stylus that meets the application. The longer the stylus, the greater the bending and deflection, and the lower the accuracy. This is not a matter of belief; it is a law of physics. The second rule is to minimize the number of joints: every extension bar or stylus added introduces another potential bending point. The third rule is to use as large a stylus tip as possible. A larger tip increases the clearance between the stylus ball and the workpiece, reducing false triggers caused by the stylus shaft contacting the part, and it also averages out the effects of surface roughness. You also need to distinguish between two length terms: total length L is measured from the mounting face at the rear of the stylus to the centre of the ball; effective working length (EWL) is measured from the ball centre to the point where the stylus shaft would contact the feature being measured. EWL is the parameter that actually determines how deep you can measure. One more point: the official guidance does not recommend using very long stylus-plus-extension-bar combinations on standard kinematic trigger probes, because rigidity drops noticeably. Only with strain-gauge probes, where trigger force is extremely low, are longer combinations permitted. So when selecting a length, first work out whether the EWL is sufficient, then apply the “shortest” principle.

Q5. We machine a lot of aluminium parts. Some people say ruby stylus tips are not suitable and we should switch to silicon nitride. Is that true? Do we need to change for all aluminium parts?

Answer: Not all aluminium parts require a different stylus. The official guidance is that replacement is only needed for “heavy-duty scanning on aluminium parts”. Renishaw’s original text is clear: ruby stylus tips are the industry standard and suit the vast majority of applications — “very few applications are unsuitable for ruby” — but two exceptions are listed. The first is heavy-duty scanning on aluminium parts, where aluminium and ruby attract each other, causing adhesive wear; aluminium transfers to the ball and builds up, so silicon nitride is recommended. The second is heavy-duty scanning on cast iron parts, which causes abrasive wear, so zirconia is recommended. Note the key term is “heavy-duty scanning”, not routine measurement. If your aluminium measurement is light and conventional, ruby is perfectly adequate. In addition, silicon nitride balls show significant abrasive wear when scanning steel, so their use should be limited to aluminium parts — they are not a universal replacement. To decide whether to change stylus material, first consider the workpiece material, then confirm whether the application is heavy-duty scanning; only replace when both conditions are met.

Q6. I bought a second-hand Zeiss CMM and previously used Renishaw styli. I now want to buy new ones — can I simply buy Renishaw? Does Zeiss have its own stylus specification?

Answer: Zeiss equipment uses a separate stylus catalogue, part number H-1000-3107 (84 pages in the Chinese edition). This does not mean Zeiss machines can only use Renishaw styli, nor that all Renishaw styli are directly compatible with Zeiss. Zeiss’s stylus interface system differs from the standard systems used by Hexagon and Renishaw, which is why a dedicated catalogue is provided. Before purchasing a stylus, first confirm the probe model installed on your machine, then refer to H-1000-3107 for the corresponding thread specification and part number. If you have an existing Renishaw stylus and want to continue using it, you must verify that its thread and length match your Zeiss probe—do not assume compatibility. Additionally, the official Renishaw catalogue contains an important note: thread specifications are determined by probe model, not by machine brand. Different Zeiss probe models may use different thread specifications. The safest approach is therefore to identify the probe model first, then check the corresponding official catalogue. If the stylus is listed, purchase accordingly; if not, rely on the official data sheet and do not make assumptions.

Summary

Stylus selection can be as complex as writing a book, but in simple terms it comes down to one sentence: first determine the stylus tip diameter, then check the stylus length, and finally verify the thread specification—this order must not be reversed.

Tomorrow, when you’re at the machine, pick up any stylus and first check the stylus tip diameter printed on it (e.g., Ø2, Ø3). Confirm it matches the minimum inner diameter and maximum hole depth of the feature you’re measuring. If the tip diameter is suitable, check the effective stylus length: use a caliper to measure from the centre of the stylus tip to the thread root, compare it with the depth of the deepest feature on your drawing, and leave sufficient allowance—refer to actual working conditions and official recommendations. Finally, unscrew the stylus and use a thread gauge or visually inspect the thread form to confirm whether it’s M2, M3, or M4. This step is the most often overlooked, yet it’s the most common cause of getting stuck when changing styli.

If all three checks pass, the stylus is generally usable. If any one fails, don’t compromise—switch to another stylus or adjust your measurement strategy (e.g., use a star stylus, extension bar, or angled head). Remember, a stylus is not a consumable; it’s part of the measurement system, just like a cutting tool is to a machining centre. A stylus with insufficient rigidity or excessive length can introduce errors larger than the machine’s own, and these errors are systematic—hard to fully eliminate through calibration.

So, next time someone asks, “What stylus do you use on your CMM?”, you can simply answer: “The one with the right tip diameter, sufficient stem length, and matching thread.”—It’s that simple, and that practical.


Need a more specific solution?

Machine configuration, operating conditions, and budgets vary from shop to shop, so this article is only a reference framework. You can:

  • Send us your machine model and machining requirements (e.g. “VCN-530C + BT40 + aluminium batch parts”) — we will match the right probe model and configuration for you.
  • Describe the probe or tool setter fault symptoms — we will give you a free preliminary assessment of the fault severity and repair direction.
  • Request a quotation or book an on-site survey — our engineers will visit, evaluate, and then issue a formal proposal.

Ningbo JCETech — 15 years focused on precision measurement, covering selection, installation, calibration, repair, and maintenance for the full range of Renishaw, Marposs, and Blum-Novotest probes and tool setters.

Contact Information Details
Official booking jcetech.cn/contact
Hotline 157-5780-7400 / 132-2194-1413
Repair shipping address Ningbo Yinzhou R&D Park A2-12B17
Service coverage Nationwide on-site service · 48-hour response for urgent repairs

Send us your problem and let our engineers help you work out the numbers.

Talk to an engineer

If the criteria or procedures in this article do not match your machine, send us the control system, probe model and a photo of the installation — we will answer with the specific settings rather than a generic recommendation.

Item Detail
Company NINGBO JCE INDUSTRIAL TECHNOLOGY CO., LTD. (JCETech)
Address Room 14-1-17, Building 033, Building 2, No. 15, Lane 587 Juxian Road, Juxian Subdistrict, Hi-Tech Zone, Ningbo 315048, Zhejiang, P.R. China
Contact Mr. Li / Mr. Zhang
Phone +86 157-5780-7400 (Mr. Li), +86 132-2194-1413 (Mr. Zhang)
Online booking jcetech.cn/contact

Repair intake and on-site service use the same address. Please call before shipping a unit so transport and scheduling do not have to be repeated.